Linked scan at constant E2/V: Difference between revisions

From Mass Spec Terms
No edit summary
No edit summary
Line 7: Line 7:


== [[Orange Book]] Entry ==
== [[Orange Book]] Entry ==
{{orange
{{orange|
This may be performed with a sector instrument incorporating at least one electric sector plus one magnetic sector. The electric sector field, E, and the accelerating voltage, V, are scanned simultaneously so as to maintain E 2/V ratio at constant value equal to the value of this ratio that transmits the main beam ions through the electric sector. The magnetic sector field is set at a fixed value such that main-beam ions of a predetermined mass/charge ratio are transmitted by the magnetic sector; these preselected main-beam ions are the precursor ions whose fragment ion spectrum is required. The fragmentation reactions thus observed occur in a field-free region traversed before the two sectors scanned in this way. The term 'E 2/V linked scan????????????????????????? is not recommmended.
This may be performed with a sector instrument incorporating at least one electric sector plus one magnetic sector. The electric sector field, E, and the accelerating voltage, V, are scanned simultaneously so as to maintain E 2/V ratio at constant value equal to the value of this ratio that transmits the main beam ions through the electric sector. The magnetic sector field is set at a fixed value such that main-beam ions of a predetermined mass/charge ratio are transmitted by the magnetic sector; these preselected main-beam ions are the precursor ions whose fragment ion spectrum is required. The fragmentation reactions thus observed occur in a field-free region traversed before the two sectors scanned in this way. The term 'E 2/V linked scan????????????????????????? is not recommmended.
}}
}}

Revision as of 22:46, 11 July 2009

Obsolete Template


Orange Book Entry

Orange Book

ORANGE BOOK DEFINITION

IUPAC. Analytical Division. Compendium of Analytical Nomenclature (the Orange Book). Definitive Rules, 1979 (see also Orange Book 2023)

Linked scan at constant E2/V

This may be performed with a sector instrument incorporating at least one electric sector plus one magnetic sector. The electric sector field, E, and the accelerating voltage, V, are scanned simultaneously so as to maintain E 2/V ratio at constant value equal to the value of this ratio that transmits the main beam ions through the electric sector. The magnetic sector field is set at a fixed value such that main-beam ions of a predetermined mass/charge ratio are transmitted by the magnetic sector; these preselected main-beam ions are the precursor ions whose fragment ion spectrum is required. The fragmentation reactions thus observed occur in a field-free region traversed before the two sectors scanned in this way. The term 'E 2/V linked scan????????????????????????? is not recommmended.

IUPAC 1997 Orange Book Chapter 12
Index of Orange Book Terms