DRAFT DEFINITION
Transmission quadrupole mass spectrometer

A mass spectrometer that consists of four parallel rods whose centers form the corners of a square and whose opposing poles are connected. The voltage applied to the rods is a superposition of a static potential and a sinusoidal radio frequency potential. The motion of an ion in the x and y dimensions is described by the Matthieu equation whose solutions show that ions in a particular m/z range can be transmitted along the z axis.

Considered between 2004 and 2006 but not included in the 2006 PAC submission
This is an unofficial draft definition presented for information and comment.

Recommended terms | Full list of terms


Related Terms