Mass-analyzed ion kinetic energy spectrometry: Difference between revisions

From Mass Spec Terms
No edit summary
Line 7: Line 7:


[[Reference:Sleno 2004]]
[[Reference:Sleno 2004]]
[[Category:M/z Separation]]
[[Category:Reactions]]

Revision as of 14:17, 21 January 2005

{{DefName|In a reverse geometry sector mass spectrometer, the measurement of the kinetic energies of the fragment ions formed in the region between the magnteic and electric sectors. A beam of [[precursor ion]s is isolated with a constant B value in the magnetic sector and the field E in the electric is scanned to determine the fragment ion masses. The width of the resulting peaks gives the kinetic energy released during fragmentation}}


References

Reference:Ervin 2001

Reference:Sleno 2004