Linked scan at constant B2/E: Difference between revisions
From Mass Spec Terms
No edit summary |
No edit summary |
||
Line 1: | Line 1: | ||
{{ | {{Def2| | ||
A [[linked scan]] performed on a [[sector mass spectrometer]] that incorporates at least one [[electric sector]] plus one [[magnetic sector]] in either order. The [[accelerating voltage]] is fixed and the magnetic field, B, and the electric field, E, are scanned simultaneously so as to maintain the ratio B<sup>2</sup>/E at a constant value. This [[linked scan]] records a [[precursor ion spectrum]] of dissociation or other reactions occurring in the [[field free region]] preceding the two sectors. The term ''B<sup>2</sup>/E linked scan'' is | A [[linked scan]] performed on a [[sector mass spectrometer]] that incorporates at least one [[electric sector]] plus one [[magnetic sector]] in either order. The [[accelerating voltage]] is fixed and the magnetic field, B, and the electric field, E, are scanned simultaneously so as to maintain the ratio B<sup>2</sup>/E at a constant value. This [[linked scan]] records a [[precursor ion spectrum]] of dissociation or other reactions occurring in the [[field free region]] preceding the two sectors. | ||
The term ''B<sup>2</sup>/E linked scan'' is deprecated. | |||
}} | }} | ||